I make ESI emitters for LC-MS nano-column packing with P2000 Sutter puller. The problem is that I cannot find a parameter set which will provide me with a large orifice for the emitter (in the manual they promise an orifice from 2 to 15 um). So far I get emitters with the cone ID around 3-5 um for 75 um ID capillaries and ~1 um for 20 um ID capillaries more or less independently of the parameters I use. I want to get 10 um for both of them. Parameters I tried so far ranged in temperatures from 240 to 330oC (no principal difference), pull force 10-20 (similarly, no principal difference) and time delay 128 everywhere. It is possible to go very far from recommended parameters, but it will take time.