From your file, it says 'iron' with a density of 3.4 g/cc. It does not sound very porous. If your sample is non-porous iron, the amount adsorbed of N2 at 77 K will be very very little. So I think you are observing the fluctuations due to experimental error of the machine.
As a thumb rule, you need to have at least a few m2 of sample inside the sample holder to obtain a proper isotherm. This will never work with a sample such as non-porous iron, for example.