We utilize Shimadzu GC-2010 GC-FIDs and using a supelco SP-2560 column. We notice that we experience a significant retention time shift with injection 1 and gradually the retention time will shift back to normal in about 10-20 injections. We mainly see this after the GC sits idle for extended period of time (>1 day). Basically every Monday at start of new work week we experience this on all of our GCs. We utilize Gas Generators for all our gases but also have gas cylinders as back up and we experience this using both type of gas supplies.
We integrate batches using Methods so this will pick the 27 peaks for us. However, with the shift the methods will not work as the parameters we set for the integration do not line up. Also, within our methods we set the retention times for the 27 peaks of interests and this shift creates issues as it will identify the wrong peaks or no peaks at all. This will create issues as we use that information to report results to clients. For the retention time shift here is the reason why it is significant. 1 picture is a sample at the start of the batch and 1 picture is the same sample at the end of a batch. The same integration method is ran through each sample. As you can see the sample at the start of the batch has no peaks integrated correctly and the identification of the peaks would be incorrect if the sample was integrated.
Example of Retention time Shift for first 15 injections of our ISTD peak.
Ret. Time (min.) Injection 1 9.951 Injection 2 9.990 Injection 3 10.003 Injection 4 10.010 Injection 5 10.016 Injection 6 10.020 Injection 7 10.022 Injection 8 10.024 Injection 9 10.026 Injection 10 10.026 Injection 11 10.028 Injection 12 10.028 Injection 13 10.029 Injection 14 10.029 Injection 15 10.029
What might be causing this to occur? How could we correct this issue?