We use the Agilent's ZORBAX Eclipse Plus C18 RRHD 2.1×50 mm 1.8u columns, now for several years.
I use K-phosphate buffer as A and methanol as B.
This particular column was taken new recently, I've run few analysis*, then washed with water and methanol thoroughly and stored. I think in the meantime nobody else used this one column. Last week I took it out and noticed the pressure with 0.05ml/min (our no-analysis mode) was about 30 MPa (usually it's something like 8-12MPa). During the analyses the pressure rose from 45 to 60MPa, but the samples were supposed to be clean (enzymes reactions I run always with purified enzyme). The peaks looked good, just the few first analyses had high background in UV (rise of baseline). Next I've run other samples, the pressure went slowly down to 55MPa and peaks still looked well (tailing f. 1.2 and width 0.08), the pressure after analyses was 20 MPa. During today's runs, the pressure was at 45MPa (since the beginning) and now is at 10 MPa, but the peaks were very broad in the beginning (0.115) with tailing f. 1.3, so almost normal. However, during the analyses the width went to 0.09 again.
I'm almost scared to run it next time, because the pressure will be probably zero.
What do you think the problem could be? Is the column bad? Or could I have done something wrong?
*I just checked and the samples were not even for me, so I cannot really check how it looked, but probably nothing suspicious.