I used two pre-cast Biorad Mini-PROTEAN TGX gradient gels (Cat No. 456-1093).
The gels were run at 140V for 60min in a Biorad Gel&Blotting chamber with a Biorad power supply.
All samples contain a similar amount of protein, so overload should not be an issue.
Already during the run, the dye front was uneven. Notably, the yellow “blurred” area appeared on both gels at the same position in the chamber, leading me to think there might be something wrong with the chamber itself, perhaps the wire is dirty or broken?
Is there a recommended way of cleaning the chamber properly in order to clean the wire without damaging it? Or do you have any other suggestions what could have gone wrong?
With the sample volume I have left I have only two more shots at this experiment before I have to start all over, so I'm thankful for any advice.
I’m attaching images of the dye front and how the lanes looked after transfer to a PVDF membrane and Ponceau staining.