Hi,

So I have been using cell titer glo 2 in the lab, but we are getting some strange results for some reason. I did some troubleshooting and found gradients across the microplate as shown below. All the wells shown here have identical contents.

This can't be due to the edge plate effect from evaporation as when running experiments with cells the outer wells are not used and are instead filled with media (but no cells). I have quantified evaporation across the plate, and it is negligible.

It also is unlikely to be a pipetting error as it corresponds to too large differences and the pipetting was done with freshly calibrated pipettes and with particular attention to good technique.

I thought one possible explanation might be the focal height (we use autofocus). I thought perhaps if the detector was too high it might be reading spillover from several wells. I ran an experiment where I filled the wells with an ATP standard and ran the luminescence assay at different focal heights. There was no discernible change in the overall pattern at different focal heights.

I thought maybe the well positioning might be off but I checked and the correct plate dimensions appear to have been selected in all experiments plus we do not see the same pattern with fluorescence readings. The platereader also has an aperture which prevents spillover while reading luminescence values and this was fitted on all occasions. This trend also remains with two different plate designs and all the plates we use for these assays have white sides so direct spillover should not be occurring. Also, all the plates are read with the lid off so any scratches/inconsistencies/marks on the lid would not affect the results.

There is also no reason to think that there is photobleaching etc. going on from the measurement as when the same plate is repeatedly measured the values in each well remain relatively constant. The temperature was also controlled and kept constant at 25C

I am a little unsure what to try/how to troubleshoot further. I may try turning the plate around and seeing if the trend persists. Possibly I might also try re-analysing the data by looking only at the centre well results (we use a 3 x 3 matrix scan) or run an experiment measuring only in the well centre. I have contacted the manufacturer as well.

Is there anything obvious I could be missing here?

More Alexis Gkantiragas's questions See All
Similar questions and discussions