I analyse SSRIs (pKa values of them between 9-10) in human plasma by HPLC following liquide-liquide phase extraction. The residue was resuspended with mobile phase (0.25 ml), water: ACN (42:58) (pH:3.5, adjusted by TFA, isocratic), vortex 15 sec., ultrasonic bed 2 min. then vortex 15 sec. The recoveries were lower than 40%. Same tubes were resuspended with 0.25 ml DMSO and re-analysed again and the additional recoveries were more than %40. So, the mobile phase does not solve all residue of analytes from the bottom of the glass tube. I think mobile phase is not always an ideal solvent for resuspension!

Is there any relation between pKa values of molecules analysed and the pH of solvent used for resuspension of residue after extraction?

Thank you