Recently I noticed that air bubbles constantly appeared from my pump and keep clogging my columns.

【Solvents】

  • Solvent A =sodium acetate buffer (0.05 M, pH 6.0) in water containing 5 mM β-CD (Sigma-Aldrich, UK)
  • Solvent B=sodium acetate buffer (0.1 M, pH 6.0 anhydrous) in a water:methanol: acetone mixture (volume ratio 20:72:8).

【degas and filter situation】

  • Because of the degasser damage, I only ultrasonic degassed my solvents for 20 mins.
  • filtered properly

【Issue】

When each solvent went through the single channel, there were no air bubbles (e.g. Flow 1.2 channel A 100%). However, when I chose 80%A+20%B, Flow 1.2, it generates tiny air bubbles. Please see the photo I took.

【potential answer】

1. solvent is still not enough degassed

2. the piston or seal in the pump was worn