We use a custom made cradle with 2 leads for ECG, respiration and temperature monitoring during cardiac MRI in our mice. We use the PC-SAM computer interface. We have been having several different problems with erratic ECG:
1. The ECG looks fine on PC-SAM but gating is erratic, missing beats etc. so that shimming takes a long time. The problem seems to be only during certain scan sequences.
2. We have a good ECG signal then slowly infuse manganese chloride through a tail vein cannula (remotely from outside the scanner room) while the mouse is still in the scanner. This causes the ECG to become so noisy that gating is not possible and the heart rate registers as being over 1000bpm. The problem often resolves itself a while after manganese infusion is ceased. We do not believe that any moisture is reaching the ECG monitoring module.
Anyone have any ideas? These problems are driving us crazy.