12 December 2015 26 347 Report

Hello everybody,

recently, I have problems with durability of my HPLC columns. Earlier, it lasted at least 3 months, sometimes up to six (that good at our department, often the columns used by others lasted much shorter time, but nobody here has any realy HPLC training). However, several last columns deteriorated very quickly and I have no idea whether I'm doing something wrong or it can be just coincindence of combination of old columns (long time unused since purchased) with old used column that just happened to collapse right now.

The problem also is, that I'm trying to detect new compounds, so I'm changing conditions and thus I'm not really able to recognize, whether the change in peak shape is due to wrong pH (close to some imaginery pKa) or because the column was fouled.

I'll try to list as many things as possible and I'd like if you could tell me, whether in your opinions any of these things can lead to such deteriorations.

I standardly use ZORBAX Eclipse PLus C18 column from Agilent. In the past I used pH 4.0, but because of some dificulties I switched to phosphate pH 7.0. During my playing, I was changeing the pH in the range about 5 - 7.5, but I don't think I would ever go higher. I operate at 0.4 ml/min and 40°C, which should be well below the limits for the column. After a batch is finished, the column is cooled down to 25°C and flow rate adjusted to 0.05 ml/min with 50%B (we were told this is better for the pumps then shutting it down). When the machine is to be turned off, I put the column into H2O/MeOH and wash usually at 0.2 - 0.3 ml/min with 5% MeOH first and then with 95% MeOH (I make 5% increments).

The mobile phase is filtered. We have in-line filter, I have changed it several times recently, but it always looked rather clean (in comparison to previous occasions). I have two types of samples - enzymatic reactions and plant extracts. The enzymatic reactions are usually stopped with double volume of methanol, centrifuged and filtered. However, because I was measuring some low concentrations recently and needed low concentration of methanol, I stopped them with heating, centrifuged and filtered. This could lead to higher load of salts, on the column, but on the other hand, the starting concentration of methanol was very low, so it should wash out, shouldn't it? The other samples are extracts from plants into organic solvents, thus they contain plenty of compounds, but those (hopefully) elute.

The first column where I noticed the quick deterioration I managed to partially dry out. I ran out of mobile phase A, but B was still there, so it was like 50% methanol, 50% air. But as I read on the Internet that shouldn't be such big concern anymore for the modern columns.

The other thing is, that I had to decrease the methanol concentration as much as I ended up with 0% in the beginning of the gradient. However, can this lead to peak spliting?

The second column was core-shell Kinetex 2.6u column we got long time ago (maybe year or two) to try it. This one lasted also only few weeks.To my surprise the main problem (at least in the beginning) was with the low flow rate, where the pressure quickly rose (between days, but also when I was washing the column with H2O/MeOH). However, when I started my analysis with the high flowrate, the pressure was about the same as the first time. I tried backflushing this column as it is possible, but that didn't help.

And last week I took ZORBAX column that was used previously many times. I checked and it worked fine. First I loaded some extracts and it worked OK (although it's hard to say with such low and so many peaks, but on fluorescence detector it seemed OK). Yesterday I run some activity measurements. The very first sample had (almost) split peaks (on attached picture). But later it actually improved, the peaks were narrower (although still wide in comparison with normal runs 0.15 vs 0.07) and tailing was 1.1, but that's rather because it's front tailing as you can see on the second picture.  So it seems rather like in the beginning there was more of the first variant, while it slowly shifted to the second variant, but not enough to split the peaks. However, even when there was less of the first compound, it did not split into two, but rather remained one wide peak (see the third picture). Nevertheless, in any case the very first peak seems to be split. Also, while the backpressure with the low florrate on this column was around 30 MPa in the previous days, now it's only around 15-20MPa, if that may be of any help.

The problem is, that most of the things I have done also previously, when the columns lasted for longer, so the main changes are the air in the column (only in the first though), old unused column (only the second though); low methanol concentration in the gradient and possibly high salt in the sample (both only in the first and second columns though).

More Tomáš Hluska's questions See All
Similar questions and discussions