its possible to have situation like this,the difference in the migration of your ladder dye front and sample as seen in your gel is not a problem because of difference in MW of the protein components in each sample and your marker.
I agree with the above answers. The dye in your samples is probably different to the dye in the molecular weight marker. Once your protein is still expected between bands visible for the marker you shouldn't have any issues. For peice of mind, check the manufacturer of the sample loading dye and the marker to see if they are the same.