Hi everyone,
I am currently using the Malvern Zetasizer ZS to measure the zeta potential of approximately 5 nm wide Au NPs capped with citrate ligands (10x diluted to about 10^12 NPs/mL), so I am expecting a very negative zeta (lower than -30 mV). When using the 1070 zeta potential cell, the results of each run were very inconsistent and the main problems that were diagnosed in the quality report were:
Poor distribution - there is typically always a high positive zeta potential peak, which I am going to guess comes from the cell wall flare
Cell wall flare - I do not visibly see any corrosion or bubble accumulation at the electrode contacts and have tried an undiluted Au NP sample (10^13 NPs/mL) or even changing to a brand new cell. The conductivity readings were about 0.08 mS/cm for 10x dilution and 0.8 mS/cm for undiluted Au NPs.
Looking for some recommendations on how else I can fine-tune this measurement!