I am trying to validate the detection of plant fungal pathogen through LAMP reaction. But I have an issue with the LAMP reaction. I am detecting the band in NTC (Control) also.

To rectify the problem, I replaced all the reagents except Bst polymerase with fresh vials. MgSO4 concentration also reduced 6 μM, 4 μM and 2 μM, temperature also verified on 65 oC, 68 oC and 72 oC.

But still same results. Please suggest, how I resolve this problem.

LAMP Reagents Volume (for 25 μl Rxn)

10x Isothermal buffer 2.5 μl

Betaine (5M) 4.5 μl

MgSO4 (50 mM) 3 μl

dNTPS (10 mM) 3.5 μl

Primer F3 (0.2 μM) 1 μl

Primer B3 (0.2 μM) 1 μl

Primer FIP (0.4 μM) 1 μl

Primer BIP (0.8 μM) 1 μl

Primer Loop-F (0.4 μM) 1 μl

Primer Loop-R (0.4 μM) 1 μl

Bst Polymerase (8U) 1 μl

Nucl. Free water 4.5 μl

DNA (10 ng) 1 μl