I am having contamination issues with my HPLC, specifically normal phase. We are running a gradient of 99.5% Hexane / 0.5 Ethyl Acetate down to 40% Hexane / 60% Ethyl acetate and the contamination peak appears 70% Hexane / 30% Ethyl Acetate.

We have tried new Hexane, new Ethyl Acetate, removing the guard column, trying a different column, flushing both channels with IPA for 2 hours, bypassing the degassing chamber and replacing the Hexane channel tubing/frit.

The only thing left I can think of removing/disassembling and cleaning the MCGV, would that be worth my time and if not, what are some other ideas that I haven't thought of?