After cleaning of the curtain plate, source, skimmer and Q0 of our 3200QTRAP (Absciex) Intensities of our Q3+ are 10 times lower than before. Q1+ intensities and peak widths are fine. We checked the calibration solutions, changed the capillary, vacuum is good, increased the detector. We have the same problem on two instruments that were cleaned at the same time. Can someone offer me some other solutions?