I would like to set up a taqman multiplex qPCR assay. Instead of using different dyes in my probes, I think a quencher can be used for 2 reporter dyes, e.g. FAM -TAMRA and JOE-TAMRA.
Thanks Hector for the answer. However, the filter in 7500/7500 fast system is not compatible with some of the quenchers (BBQ or BHQ). In addition, which platform do you use for probe synthesis, also recommend the quencher for Cy5
You can use BHQ quenchers in ABI 7500 systems. It is not required to select the quencher - you can select the option "none". It works. As Hector mentioned BHQ series quenchers are much better than TAMRA. Also check for spectral overlap between the excitation/emission of fluorophores, when you are multiplexing.