Is there any way other than prediction from signal to noise to ratio?
In case of s/n, the s/n for the peak near its base should be considered or the ratio should be calculated from the base line noise in the full HPLC run?
3)Based on the Standard Deviation of the Response and the Slope:
The detection limit (DL) may be expressed as:
DL =3.3 *σ/S
where σ = the standard deviation of the response
S = the slope of the calibration curve
The slope S may be estimated from the calibration curve of the analyte. The estimate of σ may be carried out in a variety of ways, for example
(A)Based on the Standard Deviation of the Blank: Measurement of the magnitude of analytical background response is performed by analyzing an appropriate number of blank samples and calculating the standard deviation of these responses.
(B)Based on the Calibration Curve:
A specific calibration curve should be studied using samples containing an analyte in the range of DL. The residual standard deviation of a regression line or the standard deviation of y-intercepts of regression lines may be used as the standard deviation
Same thing is applicable for the LOQ but change in point 3)
Based on the Standard Deviation of the Response and the Slope
The quantitation limit (QL) may be expressed as:
QL =10* σ/S
where σ = the standard deviation of the response
S = the slope of the calibration curve
The slope S may be estimated from the calibration curve of the analyte. The estimate of σ may be carried out in a variety of ways for example
same examples and for both
Recommended Data
The quantitation limit and the method used for determining the quantitation limit should be presented.
The limit should be subsequently validated by the analysis of a suitable number of samples known to be near or prepared at the quantitation limit.
and this is othentic document proceder as Per ICH-Q2(R1)
this is not so simple. There are three different ways of calculation of LODs and LOQs , depending on the blank result, and S/N ratio. further you can see:
Sanagi MM, Ling SL, Nasir Z, Hermawan D, Ibrahim WA, Abu Naim A. Comparison of signal-to-noise, blank determination, and linear regression methods for the estimation of detection and quantification limits for volatile organic compounds by gas chromatography. J AOAC Int. Nov-Dec;92(6):1833-8. 2009.
As per ICH guideline if your developing an assay method; there is no need of practically find out the LOD and LOQ level. It can be directly calculated from calibration curve. If you are developing method for impurity profiling you have to prepare serial dilution of your standard solution and inject in decreasing order. after that calculate the s/n ration. For LOD it should be 3:1 and for LOQ it should be 10:1.
3)Based on the Standard Deviation of the Response and the Slope:
The detection limit (DL) may be expressed as:
DL =3.3 *σ/S
where σ = the standard deviation of the response
S = the slope of the calibration curve
The slope S may be estimated from the calibration curve of the analyte. The estimate of σ may be carried out in a variety of ways, for example
(A)Based on the Standard Deviation of the Blank: Measurement of the magnitude of analytical background response is performed by analyzing an appropriate number of blank samples and calculating the standard deviation of these responses.
(B)Based on the Calibration Curve:
A specific calibration curve should be studied using samples containing an analyte in the range of DL. The residual standard deviation of a regression line or the standard deviation of y-intercepts of regression lines may be used as the standard deviation
Same thing is applicable for the LOQ but change in point 3)
Based on the Standard Deviation of the Response and the Slope
The quantitation limit (QL) may be expressed as:
QL =10* σ/S
where σ = the standard deviation of the response
S = the slope of the calibration curve
The slope S may be estimated from the calibration curve of the analyte. The estimate of σ may be carried out in a variety of ways for example
same examples and for both
Recommended Data
The quantitation limit and the method used for determining the quantitation limit should be presented.
The limit should be subsequently validated by the analysis of a suitable number of samples known to be near or prepared at the quantitation limit.
and this is othentic document proceder as Per ICH-Q2(R1)
Hello everyone! I don't really understand.. If S/N(signal to noise)=2H/h (from Ph.Eur.) and LOD(e.g.) is S/N=3, then result that LOD is 2H/h=3 and therefore, LOD is 3/2, and is not 3 or 2, but LOD=1.5H/h and LOQ is 5(not 10)H/h. Am I right? or I misunderstood something.
"Most workers calculate S/N just as the ratio of the signal to the noise, so S/N = 367/66 = 5.56 in this example. However, The United States Pharmacopoeia (USP) and European Pharmacopoeia (EP) use a different calculation where only half of the noise band is considered. In their documentation, they call the signal H and the noise h, defining S/N = 2H/h. You can see that this reported value would give S/N = (2 x 367) / 66 = 11.1 in the present case, which is twice the intuitively defined ratio."
We have developed a new approach to use metrology: it is a software platform able to organize the metrology in different application field, such as an analytical laboratory, etc. Specifically, it permits to organize the calibration of all laboratory instrumentations (e.g. scales, pipettes, thermometers, etc.) and the intermediate checks by the use of control charts, to elaborate data and estimate the uncertainty of measurement with the use of different approaches, to verify during the time the metrological characteristics of the test methods by means different tools that assure the quality. Moreover, the validation protocols of the methods for the measurements can be organized by the use of statistic tests to realize the same protocol, in accordance with the technical sectorial standard.
If you are interested in this software platform, please, contact me.