Not very much. Your MS will get a little dirty like it would with any other sample. The main issue with using phosphate buffers is that they are non-volatile and clog up the MS. The phosphate buffer will accumulate around the cone on ESI systems. If your compound has a phosphate group, it will just dirty up the system about the same as other samples you may run.
Not very much. Your MS will get a little dirty like it would with any other sample. The main issue with using phosphate buffers is that they are non-volatile and clog up the MS. The phosphate buffer will accumulate around the cone on ESI systems. If your compound has a phosphate group, it will just dirty up the system about the same as other samples you may run.
The problem is not with the phosphate but the quantity. if you use phosphate buffer then it is not volatile. as large qty. of buffer passes in ESI or APCI which is difficult to remove and large amount of salt formation takes place in MS. Thats why all types of buffers are avoided if possible. if your drug is having the posphate group then it is not the issue. bcoz total conc. of drug is in nanogram. you can use TFA instead of buffers as it volatile and acidic(0.1%).
Like everyone mentioned above, no special precautions are needed. The amount of your compound is very tiny compared to any buffer you might run. Like Rajendra stated, there is a relatively large quanitiy of the phosphate buffer compared to your sample and that makes all the difference.
If you need to run buffers in the moderate range (ie no TFA) try ammonium formate or ammonium acetate. Using a volatile buffer system will not harm the source or the MS. I've heard that TFA is hard on an ESI source, but have no experience using it. I use primarily formic acid.
If you want to avoid interference of phosphate group on curtain plate use valco valve diverter to minimize the interferences even minimal amount of solute.
It is excellent Advantage you have to make use. Like Phospholipids, you can use Negative mode, mobile (Acetonitrile : water containing 2% NH4OH), you shall get very characteristic fragments, and M-PO4, M-H2O. If apply Positive mode, same basic mobile, you shall get characteristic m/z fragments of polar positive head.
If i want to analyse water soluble vitamin in LCMS analysis Lot of article told phosphate buffer only. But not able to do that. i tried to ammonium acetate, ammonium formate, 0.5% formic acid and from 2to8 PH. then tried lot of C18 and C8 column also. But B6 vitamine elute in AQUAS 100% itself. so not able to go for gradient. please advice me some idea. In vitamin B6 elute near the void volume. All buffer and All the column, All pH. Still now we are trying.