hello, you have a beautiful pictures but, as far as I am concerned, it is necessary to reduce the volume of blood- there is some overlap between adjacent frames-
for image 1: I've Noticed That There are Some comets that do not migrate to the anode
hello, you have a beautiful pictures but, as far as I am concerned, it is necessary to reduce the volume of blood- there is some overlap between adjacent frames-
for image 1: I've Noticed That There are Some comets that do not migrate to the anode
Cells seem to overlap, therefore you can not measure the comet parameters. You should reduce the cell number. You can add PBS to 1 x 105 cells/ml and then also combine cells at 1 x 105/ml with Agarose at a ratio of 1:10 (v/v).