Hi everybody
I would like to ask why does YSI EXO2 turbidity sensor give negative values for turbidity (NTU) in seawater, even tho it's newly calibrated.
Thank you so much in advance,
Cheers
Hi.
Is the sensor intergated to a multiparameter?? if yes, just disconnect and connect again, and it shoub be working if the other parameters are ok.
Regards. Y
It could be because of dust during calibration which means you are calibrating at a value say 2 NTU and when you go to clear water like sea water, it might bring negative values.
Dear Mohanad Mushtaha,
The attached file, I hope it helps you to identify the reason.
Best regards
https://www.fondriest.com/pdf/ysi_exo_turbidity_manual.pdf
Is D-dimer elevated with tumor cells?
30 April 2024 8,349 1 View
What is the difference between 4G and 5G technology?
05 May 2023 8,690 4 View
I need a scientific definition of deep learning and its applications in areas of life
30 April 2023 4,829 3 View
A magazine printed a survey in its monthly issue and asked readers to complete it and send it back. Over 1000 readers did so.What is the type of thais sampling?
20 May 2021 3,893 19 View
Covid 19 and 20
25 December 2020 1,658 2 View
It is still debatable whether COVID-19 might recure after recovery. Is the probability of getting a re-infection still high? and is there a specific period of time after recovery where the patient...
05 December 2020 5,144 27 View
Well, both measure association between a binary outcome variable and a continuous or binary predictor variable. What is the basic difference between the odds ratio and the relative risk and when...
02 October 2020 3,457 4 View
Method and solvent
03 February 2020 7,868 2 View
Is it logical in analysis of some factor and outcome event, the results reveal significant RR less than 1 and non significant P-value?? If yes how can we interpret these results??
16 January 2020 5,437 2 View
Research Methodolgy
26 December 2019 2,104 8 View
I am trying to reconstitute the bacterium B.ovatus ATCC. it was first placed in a THIO broth and despite this we observed no turbidity. This was done 24h after the stick containing B.ovatus ATCC...
12 June 2024 1,644 3 View
Hello, Using the satellite imaginary Ive calculated the Turbidity Index of a waterbody. Is it possible to convert NDTI into Turbidity (NTU)?
03 June 2024 4,188 1 View
Hello, in the research i'm working on,I have read many references that use a silver nitrat solution in acetonitrile with Tetebutylammoniumperchlorate . however, i have been unable to obtain this...
23 May 2024 4,373 0 View
Thermal water has Na-119 ppm Cl-469 ppm and SO4 -406 But it is not located anywhere near seawater area
01 May 2024 3,836 5 View
Hello everyone, while trying to run a Visual MODFLOW Flex model I get an error information regarding the .BT3 file. Error Information: forrtl: severe (24): end-of-file during read, unit 68, file...
19 April 2024 4,172 0 View
How would acidification of seawater affect marine organisms and ocean acidification connected to global warming and climate change?
06 April 2024 1,672 2 View
Hello everyone ! I got turbidity 448 NTU in infrared light 860 nm ( ISO 7027 ) using a turbidimeter ( TB211 IR ). And it's not very clear to me exactly what formulas can be used to calculate...
29 March 2024 4,816 1 View
I am using Ag/AgCl electrode for alkaline seawater splitting. As 1M KOH is not suitable for Ag/AgCl nbecause high alkaline condition electrode potential is Ag2O. I am using 0.5M KOH. For extra...
11 March 2024 6,019 1 View
At first try, I got the cell but after that I'm not getting the cell. Can someone help me where I must be going wrong during the procedure.
04 March 2024 3,408 1 View
Hi Folks, There is an urgent need to relocate a beehive (pic attached) from an ongoing construction site in Saudi Arabia. As is evident from the pic, it happens to be on a coral bed above the...
17 February 2024 277 5 View