Hi all I could really use some inputs on why I cannot go whole-cell.

Ever since I changed the filament on our pipette puller (Sutter p87), I have not been able to get good whole-cell recordings.

I have tried several times on slices from mice of different ages, but it does not work. There is no problem getting gigaseal.

I have been doing patch clamp for a few years now, and I have never encountered this before.

Pipettes:

- pulled on Sutter p87

- 1.5 x .86 capillaries

- 2.5 mm box filament

- 4-7 MOhm - Short taper

- Under 40x magnification the tips look fine.

I used to apply .2 to .3 ml neg pressure using a 1 ml syringe to go whole cell. When I do this now, nothing happens. If I increase the neg pressure sometimes I get a very leaky whole cell with access resistance >40 MOhm and I need to inject 200-300 pA to keep the cell at -70 mV. But usually I need to apply >.5 ml before anything happens. I still get poor seals.

I have also tried different duration zaps and neg pressure in combination wit zaps. It still does not produce good whole cell recordings. I have tried several pipette holders, changed all the tubing and tried to apply neg pressure by mouth.

I am now running out of ideas. Any suggestions are very welcome!

Best Nikolaj

More Nikolaj Winther Hansen's questions See All
Similar questions and discussions