07 February 2017 7 8K Report

Questions about the Sub G1 peak in the cell cycle distribution analysis.

Significant late apoptotic positive results were obtained when the cells were assayed for apoptosis using Annexin V / PI double staining (BD Pharmingen, Oakville, ON, Canada) by using Accuri C6 flow cytometer, but the sub G1 peak was not observed at the cell cycle test by PI staining, The inconsistencies in the results of these two experiments make it difficult to explain.

My main concern is if the methods we used to test the sub G1 peak is inappropriate. Alternatively, we assume that the results is reliable, and an explain is needed.

fig 1. is about how I analyzed the data included the way I gated(I did not select cell debris because I do not think debris can be seen as "cell")  I wonder whether I have make the correct choice.

fig 2. is the result of the Annexin V / PI double staining, Annexin V+PI+ is the late apoptosis stage. the result is obviously positive. So I wonder why the sub G1 peak did not appear.

      And I I don't know if the PI staining buffer I used  is correct(did I missing some components?):

pbs:1.8ml

RNase:40μl

Triton-100:10μl

PI(10mg/ml):150μl

critric acid monohydrate:2mg

fig 3 is control(compared with treated cell(fig 1))

about fig1 vs fig3, what I wonder is that should I  select cell debris in the gate in the FSC/SSC?

More Zhe Zhang's questions See All
Similar questions and discussions