Hello,

I'm having problem with an intense dip between my runs of a HPLC method. It happens right after the first run.

When the first run finishes, the absorbance of the blank run is too high and it gets flatten at that high absorbance for two minutes while DAD is balancing. When the detector gets balanced, absorbance decreases to zero and in the next 2-3 minutes before the start of the next run absorbance goes down to -100, -120, -140mAU and more negative values till the next run starts.

This intense dip is not repeatable since starting negative values are different and because of this, subtracting sample chromatogram from the blank doesn't give a repeatable result in most cases.

This problem wasn't seen before and is happening for a month. Nothing in the method has change which is as below:

Runtime: 25min

Flow: 1mL/min

Gradient

Mobile A: Buffer of sodium dihydrgen phosphate with pH=3.00

Mobile B: 30/70 Mobile A:ACN

I'm attaching three images for the first blank run, the second blank run and the sample run.

More Alireza Shirkhani's questions See All
Similar questions and discussions