Dear All,

I am using Anti-Flag magnetic beads for immunoprecipitation experiments. I want to perform elution under the non-denaturing conditions for further use of proteins as well as to reuse the magnetic beads.

My problem is when I use 100 mM Glycine - HCl (pH 2.7), I find very low elution. Upon further addition of 2x sample buffer in the same beads, I detect an intense signal of tagged protein by western blot analysis. This suggests that elution using Glycine-HCl is not effective. For comparison, I have attached the image herewith.

I have also used Glycine-HCl with pH as low as 2 and molarity of 100 - 300 mM but still no improvement. I have figured out that the problem might be due to Flag-tagged protein as switching to Anti-Flag agarose beads also has the same problem. I know about the use of Flag peptide, but that too doesn't elute the protein completely, thereby not allowing to reuse the beads.

Kindly suggest if you have encountered a similar problem. Below are the details of my experiment, if it is required.

Beads: 30 / 50 ul

Protein: 500-750 ug

Flag Tag Expression: Endogenous (not Overexpression)

Elution: 50 / 100 ul of 100 mM Glycine-HCl (2.7)

Elution time: 10 / 15 minutes using vortex mixer at room temperature

Similar questions and discussions