If your HPLC procedure is just optimized, you need to perform the following experiments:
- linearity
- LOD and LOQ
- precision and trueness
- recovery
- stability (short and long-term)
For each point you need to perform 6 independent experiments.
For linearity are necessary at least 5 non zero points at different concentration levels.
Additionally for precision and trueness you need to analyse 6 times 3 QC levels (low, medium, and high and the concentrations need to be different respect the concentration levels used in linearity) obtained by spiking (with known amount) blank matrix.
All these procedures are deeply discussed at the following links:
If you are using the AOAC HPLC procedure you may only have to verify that 'it is appropriate in your hands for your sample'. Thus, the previous suggestion.
If standard methods have not been established, the guideline mentioned below may helpful for method validation.
1 Selection of suitable column and solvent system
· C18 column has shown good separation for this type of compounds
· The mobile phase, water-methanol with a buffer (you can use gradient/Isocratic) – gradient elution is good for better separation
2 Select suitable wavelength for quantification
First you have to run stander solutions (HMF standard are available at Sigma-Aldrich) to find out suitable wavelength for quantification (285 nm is commonly used).
It is better to use UV/Visible spectrometer or DAD detector to find out λ max for quantification
3 Linear range
Inject standards solutions (from signal to nose ration is 3:1) to higher levels to get the linear range for the calibration curve. (In accordance the guideline attached)
4 Meted of Quantification
It is beret to use External calibration procedure (using peak area or high)
5 Determination of Analytical Limits
you have to determine the Limit of detection (LOD) and The Limit of quantification (LOQ) or Minimum Detection Limit (MDL) – you can calculate those valves using the method attached.
6 Recovery study
You have to conduct recovery study (in accordance the guideline attached). You can start from estimated LOQ.
7 Confirmation test
You can use spectrum of DAD of your sample at the retention time (only for positive samples) to get some confirmation evidences. If it is possible to use LC/MS, It is much better.
It looks like the final draft of an official DIN method to replace older standards.
As usual such methods include the validation data for the method chosen, and this might serve as a starting point for your laboratory validation. The benefit of having such a standard is that once the method is accordingly established in your lab, you would only need to run the lab validation, rather going through the hassle of a basic and fundamental evaluation of the entire method.
I am afraid to say, but I cannot share this document, as copyrights heavily protect it. For this reason, I provided you with the link to the publisher site.