I'm usiing an Agilent 1100 series for NP purposes, using a quaternary pump. I have used it for a long time with n-heptane and dichloromethane as solvents in isocratic runs, using standard seals (said not to be suited for NP purpose, but I needed to be able to go above 200 bars to keep my method below 2 hours).
Due to shift in retention (decreasing retention time) I tried to dry my solvents to remove water from my system. My first attempt of drying was using molecular sieves (and keeping them in the solvents on the system to remove water added to the solvent from the atmosphere). During my attempt to troubleshoot I have tried to dry by having activated silica at the bottom of my solvent flasks. It all worked rather well for some time, but suddenly I saw great drops of pressure but with different behaviors.
Scenario A) a sharp drop in pressure from 140 bars to below 100 bars, then a large fluctuation and slow build up to 140 again, where it would have larger-than-normal fluctuations (normal was 0.8 bar, new was 2-5 bars) before a new drop Scenario B) a sharp drop in pressure from 140 bars followed by fluctuation at lower pressure and a slow build up to 100 bars where it stays with larger-than-normal fluctuations before a new drop Scenario C) a slower drop in pressure followed by a prolonged time at the lower pressure-plateau with larger-than-normal fluctuations and then a slow build up of pressure
We suspected air caused by evaporation of the solvents in the filters on the lines caused by a localized vacuum if dust from the sieves had clogged the filter. Therefor the filters were exchanged to new ones, but the behavior was the same and the pump started to make noises as if I had a buffer in one of the lines precipitating. I have ended up changing every filter and seal possible in the pump as well as a seal in the ALS. I tried just connecting a long tubing for testing to eliminate the possibility of the problem being somewhere in the system following the pump. I still saw the weird behavior in pressure and by letting air into the line I could see how the pump seemed not to take in solvent (the air bubbles were not mowing) when the pressure dropped, so it is not due to a leak but rather a missing uptake. I have now tried to take a binary pump from one of our other 1100 systems, and saw the exact same behavior as soon as I put pressure on! It worked without any problems in RP-mode a few weeks back.
Both pumps can easily run at 4 mL/min when purging on all lines and with dried n-heptane, dried DCM and MeOH. But as soon as I close the purge valve (at lower flow rates) there seems to be an issue for both pumps.
I hope some of you have experienced something similar and solved it or have an idea for me to try out.