We have an old HPLC composed of a Waters 600E pump and a 410 differential refraction index detector (with built in IEEE488).

We got it a PC with a BUS LAC/E card (half width) and the millenium 32 software.

I got the pump and detector recognized by the equipment, but in the software i couldnt find a way to recognize an external trigger.

The manual of millenium says: "Detectors and system controllers that are not controlled by a Millennium32 system (for example, 2690/2695, 600 Series controller, and GC) require an inject start trigger signal (contact closure) from the injector as each injection occurs. The inject start trigger signal instructs detectors and system controllers to acquire data or run methods."

"Note: Any instrument that is not controlled by the Millennium32 software requires an inject start trigger wire connection (see Section 2.4.4, Injectors)."

2.4.4 says that for the 410 detector, the trigger (contact fromt he manual injector) must be done in the "Chart Mark" and "Dgnd" connector on the back.

I did those connections. When I inject, both the 600E pump and the 410 detector recognizes the injection, but milleniums stays at the "wating for injection" status, until one press the button on the software.

Is it possible what i want? If it is, is it possible with just the 410 and the 600E pump?

I have a 486 UV detector too and a buSAT/in box if it can be done through them (but the idea is just with the 600E and the 410 by themselves).

More Patricio Nicolás Romero-Hasler's questions See All
Similar questions and discussions