I have a problem with my Sephacryl S-300 HR column. I recently purchased the resin to pack with a 70-cm long glass column (homemade), with a 1.75 cm diameter. The bed volume is about 150 ml and the column was packed with a peristaltic pump using the flow rate described in the insert. The problem is when I tried to run a 2-ml mixture between blue dextran (BD) and potassium dichromate (PD) (3 mg/ml each and ran at 0.58 ml/min), both BD and PD seemed to diffuse throughout the column as they traveled to the end of the column. They don't seem to separate well. Something must be wrong here. BD should leave the column way before PD but that wasn't the case here. Could anyone help me troubleshooting this problem, please?  I have these questions below:

1. Does Sephacryl resin require a buffer to pack and run? My sample is quite stable in water and this is why I use it instead of a buffer. And so the packing and running were done using milli-q water. Would this be the cause of the problem?

2. I don't think the resin was bad since I just bought it from the company and I don't think BD and PD was the culprit either. I am also sure that the loading technique was not the problem either. I think the problem is the column was not properly packed. Could anyone recommend me about packing and dealing with Sephacryl resin?

Thank you very much in advance

Similar questions and discussions