I use a glass cuvette for DLS analysis (and made the necessary specification in the protocol on the Zetasizer software to select the cuvette type/manufacturer). Ever since I started using this cuvette (it's wasn't new), I was getting huge standard deviations in size measurements (~20-30%) and very low count rate (~200 kcps). I even worked under various concentrations (1-10 mg/mL) to ensure this wasn't a concentration-dependent issue.

I randomly tried using a disposable polystyrene cuvette yesterday with the same protocol and concentration and found a much higher count rate (2000 kcps) and lower standard deviation (within 10%) for the same exact sample. I now know this isn't a sample or instrument problem but a cuvette issue.

Are disposable or polystyrene cuvettes better for DLS use? Or is there an issue with my cleaning protocol (see next paragraph)? If there is something such as a sub-visible stain or scratch, how can this be determined?

Protocol for cleaning polymer sample:

1.) Rinse 3 times with DI water and dispose.

2.) Rinse 3 times with ethanol.

3.) Wet a kim wipe and wet the outer surface of the cuvette.

4.) Air dry the cuvette using the laboratory air supply. Once dry, load sample.

More Rahul Upadhya's questions See All
Similar questions and discussions