I use a glass cuvette for DLS analysis (and made the necessary specification in the protocol on the Zetasizer software to select the cuvette type/manufacturer). Ever since I started using this cuvette (it's wasn't new), I was getting huge standard deviations in size measurements (~20-30%) and very low count rate (~200 kcps). I even worked under various concentrations (1-10 mg/mL) to ensure this wasn't a concentration-dependent issue.
I randomly tried using a disposable polystyrene cuvette yesterday with the same protocol and concentration and found a much higher count rate (2000 kcps) and lower standard deviation (within 10%) for the same exact sample. I now know this isn't a sample or instrument problem but a cuvette issue.
Are disposable or polystyrene cuvettes better for DLS use? Or is there an issue with my cleaning protocol (see next paragraph)? If there is something such as a sub-visible stain or scratch, how can this be determined?
Protocol for cleaning polymer sample:
1.) Rinse 3 times with DI water and dispose.
2.) Rinse 3 times with ethanol.
3.) Wet a kim wipe and wet the outer surface of the cuvette.
4.) Air dry the cuvette using the laboratory air supply. Once dry, load sample.