We are performing a microfluidics based immunoassay reaction, with a modified system. We have been getting lower readings for our spiked samples, compared to control samples. For example, if control is giving a 0.3, then sample will give 0.2, and often the value decreases further with increasing sample concentrations. We had done the assay previously multiple times, where the readings were perfect, and this is a comparatively recent phenomenon, that we have been unable to troubleshoot.

Antibodies used are monoclonal antibodies. We are binding them to the surface using a glutaraldehyde chemistry.

We are not using any HRP or other enzymatic reaction, but a proprietary detection system.

Samples used are Thyroid stimulation hormone spiked in assay buffers.

We have analysed multiple assay buffers, starting from PBS, PBS with BSA, FBS, calibrator that comes with the commercial ELISA kit, and a few others. Trend has been similar.

Blocking buffers has been mostly protein based (BSA, Casein, FBS sera, at times) with variations including Tween-20 and few other small molecules.

We are unable to understand the cause of this, and hence are asking what is the possible reason.

More Rohan Aggarwal's questions See All
Similar questions and discussions