05 November 2023 2 7K Report

Recently, I cannot use thin-film hydration method to successfully prepare my DPPC MLV. I am sure I didn't change any parameters or steps in my protocol. However. instead of acquiring thin-film, I got transparent aggregation (see attached picture) at the bottom of round bottom beaker after rotavap overnight. The lipid and chloroform I used were purchased two weeks ago so I don't think it is due to the outdated chemicals. I ever doubt rotovaper problem since I found that I could easily pull out the connected beaker even at very high vacuum degree (6mbar). Then I connected a pressure gauge to the rotovaper to test if the rotovaper works well. I tested twice and the gauge pressure very corresponded to the value read from the rotovaper screen. Now, I am so confused, could someone give me any possible explanations about why? Many thanks!

More Kangdi Sun's questions See All
Similar questions and discussions