Dear all. I have a problem when I want to build an optical transmission microscope with a halogen light source. The design is shown in figure 1.
The basic idea is a light from a halogen light source is used as illumination. Then an objectice with 4 times magnification and 0.1 NA is used as an condenser before the sample and another objective (50 times NA=0.55). Afterwards, a 2d slit is put on the intermediate image plane to select the area of interest. In the end a bean spillter is used to split the light into the Camera and Spectrometer to get both the image and spectra of the area that we are interesting.
However, when I start to buide the system, something strange to me is happened. The beam in Plane A (which marked by red in Figure 1) is not colimated: The size of the beam is continue increase fast, and I even can see a sharp image before the later lens. What is happened for the setup. Is it due to the magnificaiton or NA of the two objectives are not camparable?
Thanks for you help in advanced.