I apologize in advance for the length of this post but there is a lot of information. Our lab has an Acquity TQD coupled with an H-Class UPLC (Column manager-CM, sample manager-SM, and quaternary solvent manager-QSM). The column I'm using is a UPLC Phenyl Hexyl 2.5um 2.1x150mm Column.

We acquired the UPLC setup from another lab that did not take care of it. It was throwing overpressure errors constantly (>14,000psi) when trying to run anything other than water. I eventually got that figured out by replacing injector valves, the vespel seal (QSM), running an acid/base wash through the system, and sonicating everything that could be sonicated.

It seems like I fix one thing and then the next problem shows itself. After all of the above, the pressure dropped dramatically (~1000psi). According to Waters that was too low. Then we were seeing problems with inaccurate flow rate volume, i.e. 0.4ml/min x 5 min =