Hello,
I am working on the installation of an Agilent ICP-MS 7700s equipment. I am only using He mode. I have tested No Gas mode and everything seems just fine. Also several CRM have been tested with acceptable results. However, when He mode is tested calibration curves and results don't make sense (calibration curves attached). Increasing the He flow seems to improve the problem in the so-called HEHe mode. The tune parameters used for each mode and some calibration curves corresponding to the three modes are attached to this question. I would be grateful if you can give me some advice about this issue.