Hi everyone,
I am trying to develop a LC/MS method for the analysis of 2 anionic surfactants. While the method seems to be working fine, I am facing an odd carryover issue.
- We are using a Xevo-TQS micro with an FTN autosampler.
- Mobile phase is 90/10/0.15 % ACN/H2O/FA with a flow of 0.2 ml/min, operated in negative mode, for 10 minutes. Peak elutes around 5 minutes.
- Needle wash is 80/20 ACN/H2O
- Tried 3 different columns and oddly, when we first start injections, we do not see carryover. But the following day, we see carryover in the blank injections. What could this mean?
- The area of the carryover does not drop with subsequent injections.
- If you inject a vial with just air in it (no solvent), the carryover disappears
TIA