I am currently developing a LCMS method for certain genotoxic impurity in API on a shimadzu 9030 qTOF LCMS. And I had an argument with my coworker about which mode is more sensitive, SIM or MRM?

My coworker said that MRM is more sensitive based on his past experience using LCMS for API analysis.

I am relatively new to LCMS. But the way I see it, MRM is definitely more selective. But it is only more sensitive than SIM when there is background interference.

MRM is a MS/MS method, which should have fewer ions compared to SIM, which is a MS method. So if we have a clean standard or a sample which has no signal at this particular m/z around the retention time of the target, SIM should be the more sensitive one.

Am I wrong?

Thanks in advance for any comments!