We have checked the injector, ECD for any contamination. However the engineer says both are alright. I am currently using an Equity 5 column.
Normally it is sign for high contamination of detector which is usually indicate the high concentration of sample has been injected.
always use very dilute solutions for GC.
first remove the detector end of column and condition the column according to the leaflet present in the box of column.
then connect the column to the detector and set the temperature of detector at 10C below it T-max with carrier gas flow for at least 1hrs.
then check the condition with dilute injection of chloroform.
if not better repeat the procedure.
I had similar problem and I decided to reset all the system, it was a problem with the communication between the computer and the GC. Clean the temporary memory in the computer and OK.
other question it is if your sample has water and if yo are using a FID detector
In most cases negative peaks in GC with ECD may occur after a positive peak due to an dirty or old ECD cell.
Follow the guidelines in the ECD manual for cleaning.
Another source are small leaks in the GC system. Try a sensitive leak detector - NOT liquids as Snoop or soap.
After removing even very small leaks, condition the detector for some hours at the max. temperature (for Ni-63 appr. 340 °C.
Can quartz wool adsorb analytes in a GC FID analysis? The reproducibility of our samples in consecutive injections is not good. Can wool be the cause?
02 March 2013 9,637 12 View
Does anyone have experience analysing B12 from food samples using HPLC-Vis?
11 December 2012 7,207 13 View
A crude extract of fungal culture using EtOH was subjected to column and TLC and partially purified compound was obtained. UV vis spectrum of the compound/s has max absorbance at 218nm. The...
11 August 2024 9,801 2 View
Can anyone explain this method? Especially the last statement where it says only at 1.5 to 2.5mins was the MS/MS connected to the UPLC. How is that possible, is it a feature in this specific...
11 August 2024 8,141 3 View
Hello experts, Does anyone know any free software about retention index prediction ?
08 August 2024 7,403 2 View
I'm working on selecting antibodies against a recombinant protein that has a His-tag. My idea is to first bind the recombinant protein to a HisTRAP column and then use this column for an affinity...
07 August 2024 505 3 View
Why is TLC not a form of partition chromatography and paper chromatography not a form of absorption chromatography?
03 August 2024 7,035 3 View
or we can collaborate
01 August 2024 6,705 2 View
I need something like Camag's VisionCat software for photodocumentation of TLCs. Could you recommend something?
31 July 2024 8,415 1 View
Hi everyone, I'd like to better separate two close peaks coming from a bioconversion sample, I'm currently using a C18 KromaPhase 4,6mm Ø.I. SCHARLAU, particle size (µm): 10, pore size (Å): 100,...
30 July 2024 435 3 View
I am working on algal extract to which gas chromatography (Not GC-MS) spectrum I want to discover. My question is can we identify specific compounds using retention time if I compared the RT with...
29 July 2024 8,034 4 View
I'm having problems with ion exchange chromatography. After applying my sample (a fluorescent protein), a preferential path inevitably forms. The matrix I use is DEAE-Sephacel. Could someone...
27 July 2024 6,605 0 View