I have information regarding mass spectrum, FTIR, UV spec maximum absorption of a microbial secondary metabolite. Is it possible to elucidate the compound using these data?
May be difficult. MS will give mol no.,FTIR will give functional group, UV, delta max, but needed to know C & H present but if you have expected compound, try merck index online.
Hi, If you have data of HR MS (ESI + or ESI - for example), I think you can predict the exact MW (confirmation with some adducts), predict the molecular formula (by isotope pattern), if you have MS and MS MS , you can predict your compounds (by using some MS data base on line such as Metfrag, Metlin, MassBank or others; For confirmation I think you need have standard or isolating the compounds as described by Jean.
No No, it is just not possible at all. Both of these instrumental techniques will add to your analysis qualitatively only . For the Quantitative thorough structural elucidations / investigations , you must go to the basic organic chemistry instrumental methods. There are many of these , more you make use of these techniques, more closer & thorough will be the OUTCOME . Dr Abid Aziz Ajaz
Complete structural identification of an unkown secondary metabolite can only be achieved using High Resolution 1D and 2D NMR..... and like JRG above, pointS out, the metabolite must be stable otherwise you will elucidate the structure of an artifact of the actual metabolite ...... NMR can then be followed by MS, UV and IR as confirmatory techniques