The negative peak always appear in the blanks.

I have tried several approaches which didn’t seem to work.

I have cleaned the HPLC with isopropanol + acetonitrile + hexane + dichloromethane (50:25:15:10%, v/v) overnight at a flow rate of 0.5ml/min without the column.

I have also injected the same mobile phase as the blank but the negative peak is still there.

I have set the reference wavelength of detection at 190 - 800nm, 200 - 600nm, 200 - 500nm and 220 - 600nm but no difference.

The last thing I did was washing the column with 100% acetonitrile but it didn’t solve the problem of the negative peak.

The column that I am using is a Waters Spherisorb 3µm ODS2, 4.6 × 250 mm.

Your help will be fully appreciated.

Thanks

More Charles Ojobor's questions See All
Similar questions and discussions