I have been working to develop a microfludic protein analysis system and the current prototype system utilizes eppendorf-type vials to contain samples and protein reagents. The vials are connected to the analysis device via peek tubing, so that the fluid flows from the vials to the device. The peek tubing is re-usable and thus needs to be cleaned or flushed between assay runs.
We typically block the tubing and device prior to the assay, but after the assay we see a significant amount of residual protein analyte that affects the next assay run. The assay is a protein binding sandwich assay.
What have we tried?
-wash buffers containing Tween20
-high pH buffers with 100mM or more NaOH
-low pH buffers with 100mM or more HCl
-isopropanol, methanol, ethanol.
-combinations of the above with and without soaking times.
Results?
-the alcohols seem to do the most, but still not a one-step solution. We would like a method that works every time, a single washing reagent would be nice. We start the system dry, so even a gas based cleaning procedure might work. I may not necessarily need to remove the protein, but rather effectively denature the proteins from the previous assay run.