Hello dear RG community.

I have a Dantec PIV setup consisting of, in particular, an iNanoSense camera and a SoloPIV 120 NewWave Research laser.

I'm getting the first image less bright and nonuniformly lit as oppose to the second image in a PIV pair of images.

First thing I did is I tried to determine whether it is the laser's or the camera's fault. I swapped the laser heads and, still, got the first image less bright and nonuniformly lit.

Hence, I made a conclusion that it is the camera's fault.

The only fix I could think of was to play with the timing diagram to try and make the camera getting the first image right. I didn't manage to come up with such a timing diagram.

Now, I'm out of ideas.

I'm wondering if anybody had the same issue, came up with a working fix and could share it with me, please.

Attached, is my timing diagram.

Thank you in advance.

Ivan

P.S. I did talk to Dantec's support to no avail. Dantec did help me a lot overall, but we couldn't solve this particular issue.

P.P.S. I don't want to play with the attenuators to try and get the first laser head more powerful to compensate for the camera's error (if that's possible at all ...).