Hi,

we have "inherited" an older UV-Vis spectrometer - a Cary 100 Bio - which is still available now from Agilent.

As no one is experienced with this device i read available information i could find on how to properly use it.

But my "problem" is that the infos are not coherent.

My main question is how to measure properly in practice.

In the manual it is not described clearly - e.g.

for the scan program it is asked to add a blank/reference when using baseline correction, while for the zero nothing is told.

While in the simple reads program it says "Make sure that the sample compartment is clear (or you have a blank in position) and zero the instrument by pressing the Zero button"

In simple reads you have no options - but the readings differ when you add/dont add some sample in the read beam. Which tells me: simple reads might be using double beam. On the other hand the simple reads manual you can zero with a clear compartment - which would make the sample measurement with sample front and blank in back look strange.

On the webpage you find (https://www.agilent.com/en/support/why-do-we-perform-a-zero-baseline-in-scan-faq) that zeroing should be done "using the actual solution (without the sample) in both the Front Beam (Sample) and the Rear Beam (Reference)."

This practice also is used in an introductionary video by supreme science:

https://www.youtube.com/watch?v=nr3uyW1_QbM

But in a video from Yale a blank is only added in the front beam holder

https://www.youtube.com/watch?v=ivDVqj6IM3s

To me it makes the most sense to always use:

- blank 1 in the front, blank 2 in the back beam compartment to zero/baseline -> so Lambert-Beer I/I 0 is the same values

- sample in the front, blank1/2 in the back for the actual measurement. No matter what the program is you use (scan, simple reads etc.)

(a procedure that is nicely described here : https://www.youtube.com/watch?v=wxrAELeXlek)

So my question is to anyone experienced with the Cary 100:

As the info i found ranges from:

Zeroing by:

- no cuvette in beam

- cuvette only in front beam

- blanks in both front and back beam

What is the correct/most accurate practice to use the device

a) for zeroing/baseline

and

b) for measuring the sample

And

- are there differences for scan programm and simple reads / other programs like kinetics, RNA/DNA

In scan you have the option to choose double beam, in simple reads you dont

The differing infos makes me unsure what procedure will give the most accurate results.

More Daryl Goodrich's questions See All
Similar questions and discussions