Dear ResearchGate Community,

I am currently facing a challenge in the LC-MS analysis of a highly hydrophobic peptide (already purified to >98% by a manufacturer) on our 1260 LC-MS System.

The experimental setup involved running a gradient of 5-80% ACN over 30 minutes on an analytical column (C18, 95Å, 4.6 x 50 mm, 5 µm). I injected 1, 2, 5, and 50 µl from a sample with a concentration of approximately 0.5 µg/µl (prepared in either 100% Methanol or 80% ACN+20%H2O+0.1%FA).

However, I observed multiple peaks at retention times between 21-24 minutes with my expected m/z in the Total Ion Current (TIC) MS. Interestingly, there were no UV peaks at either 220 or 280 nm (see pic).

Has anyone else observed such behavior (i.e., broad peaks in MS with no UV signal) with any highly hydrophobic peptide or compound?

I would greatly appreciate any suggestions or tips to resolve this issue and achieve a single peak in both TIC and UV (220nm) for this highly hydrophobic peptide.

P.S.: I can observe a single peak in both MS (TIC) and UV (220nm) with my water-soluble hydrophilic peptide, indicating that the LC-MS System is functioning properly with the chosen mobile phases (A: 0.1% TFA/water, B: 0.1% TFA).

Thank you in advance for your assistance.

More Rajesh Kumar's questions See All
Similar questions and discussions