Greetings. I would like to get opinions from the experts for the improvement and optimization of multiple drug confirmation by LC-MS/MS.

1. One of our method calibration curves follows the sample extraction protocol. The spiking material for the calibration curve was prepared at 0-1000 ng/mL in the urine matrix before extraction. Along the extraction process, the cal/QC/sample is diluted for 10x adding various reagents. However, the curve was projected for 0-1000 ng/mL when the actual amount is 0-100 ng/mL. Is my observation correct? I proposed to adjust the concentration of calibrators to have a final concentration of 0-1000 ng/mL to increase the sensitivity and accuracy of ESI positive mode. Does it dilution trade-off the error since all the samples were prepared in the same manner?

2. The standards used for the calibration curve are parental compounds. Is it necessary to have a 100% matrix match since the addition of beta-glucuronidase will not result in any reaction in parental standard materials?

I would appreciate the input from the experts. Please feel free to comment.

Similar questions and discussions