The Hach Lange kit goes from 0-1000mg/L.
Hach Lange say their LOD is around 6ppm and the LOQ is 21ppm.
Does anyone know how they will determine the LOD and LOQ for these kits?
I understand that one would determine 3xstd(blank) to get the LOD, however because these are test vials, you need to blank initially with the blank you made. Would I then just read the blank after blanking with it?
A problem with these vials is that if I do blank with the blank and read it back it gives me zero as it should. But when you move the test vial slightly in the spec, the values no longer reads zero. this is because the glass vial is not optically the same at all points (and it is round which cause light scattering). I cant just keep moving the vial around and taking any reading I want.