I attached two GC chromatogram that one of them is the primary chromatogram that we obtain and other one is the same sample and same sample preparation after 1 month! why we have this tailing in second chromatogram?? we trim the column and do conditioning about 24 hour in 210 C as well but it didn’t make a differences! Is it possible did the column's destruction happen because of tempreture or impurity of carrier gas or something else in such a short time?

Column description: Stabilwax (Restek) , 60 m × 0.25 mm , 0.5 um, temp range :40 C to 260 C

CAT # 10641 , serial No. 120214

Inj. Vol : 1ul spilit (spilit ratio 20:1)

Ing. Temp : 250

Oven temp : 110 C to 210 at 5 C/min (hold 5 min)

Carrier gas : N2

Linear velocity : 31 cm/sec

thanks in advance for your help