I would be grateful if someone would advise or give a link on a detailed fluorescence titration technique. I am interested in setting up an experiment on binding of a receptor to a fluorescein-labeled peptide ligand, choosing a control, etc.
Adam B Shapiro We have a purified receptor and fluorescein labeled peptide as ligand. Device for simple flu-intensity measurement Fluoroscan 3. We placed peptide solution with defined concentration in cuvette and then made up titration with receptor solution. We calculated Kd but it is possible we observed the receptor-fluorescein binding instead of peptide-receptor binding. That is the question...
The way to determine if the signal was due to binding of fluorescein rather than peptide is to try to compete the fluorescein-peptide off the receptor using the unlabeled peptide.