We are trying to standardize a method for analysis of TAGs using Reverse phase C18 column and RI detector in an Agilent HPLC using literature based reports. Since RI detector in our HPLC is not working below 35oC the detector and column temperature was set at 35oC. As per literature Acetonitrile: acetone, we are using as mobile phase. The problem we are facing is at ratio of 1:1 acetonitrile: acetone we are not able to achieve a stable base line despite of running the system at prolong low flow rates, daily column washing by changing the polarity gradient of both solvents, changing the flow rates and occasional detector purging. When we tried to run with different mobile phase ratios like 70:30 and 60:40 or 55:45 Acetonitrile: acetone (when concentration of acetonitrile was more) the baseline we achieved was abnormally and unbelievably straight and 100% noiseless. But when we tried to run the samples (TAG Standard) in those conditions not a single peak nor a single noise appeared which indicated that samples/analytes are probably not getting detected somehow. Hence we had to change the mobile phase ratio again into 50:50 Acetonitrile: Acetone because most of the reports suggested either 1:1 ratio of both or acetone on the higher side for TAG analysis in RP-HPLC. Also more polar acetonitrile at higher ratio was although supporting correct baseline, but was not supporting analyte detection. However, at 1:1 ratio we again lost the baseline stability (extremely noisy baseline). Can anybody suggest suitable conditions in terms of mobile phase or anything else for TAG analysis with C18 column and RI detector? Someone's kind help will be extremely appreciated.

Similar questions and discussions