Dear experts and colleagues,

I have an lateral flow assay that previously worked but this time the test line did not appear.

1) I have tried different vials of Antibody for test line and different vials of Antigen as the deterioration of protein may happen but nothing changed.

2) I check the reaction between Antigen and antibody in conjugation by dotting antigen on membrane and use conjugation as sample. It worked well.

3) I made another strip and dotted the Antibody on the membrane as test line (no control line). Then, I mixed the Antigen with conjugation and put the strip into this mixture. The dot appeared.

From above experiments, I assumed that the test line did not appear in full strip (having both test line and control line) experiment is due to the the Antigen did not react with the conjugation on the conjugation pad. So, after going conjugation pad, the sample contained Antigen and conjugation instead of Antigen+conjugation.

Have you experience the same situation? Could you please give me some advices and suggestion. Thank you for your help.

For your information: The conjugation pad was treated with sucrose, trehalose, tween 20%, BSA. The membrane was not treated.