We have recently replaced the tip seals on our scroll pump and since it came back our scroll pump has started acting oddly where the backing line settles at about 1E-1 mbar where previously it was at 3E-3 mbar. What is odd is as the main system pressure reaches the HV/UHV border the backing pressure increases. Furthermore, when the main system is flushed with a gas like Argon or nitrogen for 10-15mins. The main system returns to its previous pressure afterwards as expected but the forevacuum pressure drops into the low E-2's or high E-3's. Basically lower than it was before. This does not last though and after about 5 hours it has returned to about 1E-1 mbar where it settles.
Has anyone else encountered this before? I have spoken to the manufacturer but they are also a little confused.